X-Rite eXact Standard Edition Spectrophotometer
X-Rite eXact Standard Edition Spectrophotometer
X-Rite eXact is a next generation handheld color measurement solution that takes into account the new business needs and dynamics of a global supply chain. X-Rite eXact enables printers and packaging converters to truly understand, control and manage color across the entire color network, ensuring the utmost in color accuracy.
Exact - Standard Edition
Exact - Standard Edition
X-Rite eXact Standard: A spectrophotometer that ensures accurate printing of both CMYK and spot colors as well as leveraging industry standard for process control.
Support for All Measuring Conditions
Using a simple process, eXact fully supports all measuring conditions with one platform, measuring M0, M2 and M3 with a single pass, saving the time and aggravation of measuring each individually.
Measurement Conditions and Filters
M0: No Filter, UV included
M1: Part 1, first in the industry with D50 (Daylight) through the entire visible spectrum
M2: UV Cutoff Filter, UV excluded
M3: Polarization Filter

Measurement Conditions and Filters
M0: No Filter, UV included
M1: Part 1, first in the industry with D50 (Daylight) through the entire visible spectrum
M2: UV Cutoff Filter, UV excluded
M3: Polarization Filter

| Description |
eXact Density |
eXact Standard |
eXact Advanced | ||
| CMYK, Spot | CMYK only | CMYK and spot | CMYK and spot | ||
| Functions | Special | BestMatch | x | x | |
| Densitometry | x | x | x | ||
| Colorimetry | CIE L*a*b* CIE L*C*h | x | x | ||
| dE Trend | x | ||||
| CIE XYZ & Yxy | x | ||||
| Special | Paper Indices | x | |||
| Metamerism | x | ||||
| Absolute & Relative Color Strength | x | ||||
| Opacity | x | ||||
| Printing Plate | x | x | x | ||
| Reflection Graph | x | ||||
| Quality Control | x | x | |||
| Features | Scanning | ||||
| Sample Storage | x | ||||
| Multiple User Profiles | x | ||||
| Security Settings | x | x | x | ||
| Pantone PMS+ | x | x | |||
| Averaging | x | x | x | ||
| ISO 13655 | M0 (No-Filter, UV included) | x | x | x | |
| Measurement Conditions | M1 (Daylight, D50) | x | x | x | |
| M2 (UV Cutoff Filter, UV excluded) | x | x | x | ||
| M3 (Polarization Filter) | x | x | x | ||
| Single Pass Measurements | M0, M2, M3 | x | x | x | |
| E Method | dE*76, dE*94, dE*00, dE CMC | x | x | ||
| Density White Base | Absolute, Paper | x | x | x | |
| Interface | USB | x | x | x | |
| Bluetooth* (based on regions) | x | ||||
| Software | eXact Manager | x | x | x | |
| DataCatcher | x | x | x | ||
| Color iQC Print | |||||
| InkFormulation | |||||
| NetProfiler 3 | |||||
| PantoneLIVE |
O = Optional
Microsoft Windows XP 32 bit (with latest Service Packs and updates installed)
Microsoft Windows Vista 32 or 64 bit (with latest Service Packs and updates installed)
Microsoft Windows 7 32 or 64 bit (with latest Service Packs and updates installed)
SPECTRAL ENGINE
Spectral analyzer: DRS spectral engine
Spectral range: 400 nm - 700 nm with 10mm interval
OPTICS
Measurement geometry: 45/0 ring illumination optics,ISO 5-4:2009(E)
Measurement aperture: 1.5 mm, 2 mm, 4 mm or 6 mm
Light source: Gas filled tungsten (illuminant type A) and UV LED
REFLECTANCE MEASUREMENT
Measurement Conditions: According to ISO 13655:2009
Power supply: _X-Rite p/n: SE30-177100-240VAC 50/60Hz, 12VDC @ 2.5A
Battery: _X-Rite p/n: SE15-44Lithium Ion, 7.4VDC, 2200mAh
Physical dimensions
Dimensions: H: 3.0 (7.6 cm), W: 3.1 (7.8 cm), L: 7.1 (18 cm)
Weight: 1.6 lbs. (0.7 kg)
Accessories Provided
USB cable, Power Supply, CD with software and documentation, Docking/Charging Station, Quick Start Guide, Carrying case
Battery: _X-Rite p/n: SE15-44Lithium Ion, 7.4VDC, 2200mAh
Physical dimensions
Dimensions: H: 3.0 (7.6 cm), W: 3.1 (7.8 cm), L: 7.1 (18 cm)
Weight: 1.6 lbs. (0.7 kg)
Accessories Provided
USB cable, Power Supply, CD with software and documentation, Docking/Charging Station, Quick Start Guide, Carrying case
- PRE : NONE
- NEXT : CR-10 Plus Colorimeter with Good Price